What are the requirements for a pendulum tube rain test device?
The swing pipe rain test device is also called ipx34 swing pipe rain test chamber, ipx34 rain test device, etc. It can be divided into three types of instruments with different appearances: open type, shower device and box type.
Recently, I saw that some netizens wanted to know more about the pipe rainwater test device, such as what are its requirements? What is the testing method? Wait for a series of questions and then analyze them as follows:
First, the technical requirements and parameter requirements of the pendulum tube rain test device should comply with IEC60529 and shell protection level (IP code) GB4208-2017.
Second, the test conditions require that the product be tested by swinging the swing tube back and forth. The angle of the ipx3 swing tube is 120 degrees. º;ipx4 is 350 degrees, and the swing speed of the swing tube is 120. ° back and forth 4S; 350° back and forth 12S, test standard is 10min. The test time can also be adjusted according to the user's testing needs.
Then there is the material. The normal appearance material is made of domestic steel plates. The inner box and turntable are made of SUS304# stainless steel plates. The surface is painted, which is beautiful and durable.
In addition, its nozzle is manufactured according to IEC60529 specifications.
To sum up, here are the requirements for the pendulum tube rain test device.