Two-chamber cold and thermal shock test chamber why oil blockage will occur

Date: Oct 09,2022 Views: 292

If the operation in the two-chamber cold thermal shock test chamber tilt angle is relatively large, then the test chamber compressor refrigerant oil will flow from the suction tube into the low pressure chamber. When restarting the test chamber equipment compressor refrigeration, the refrigerant oil will be sucked into the cylinder, which can make the compressor load in a short period of time (the compression ratio in the cylinder is based on the volume of gas to determine, however, the liquid is not easy to be compressed), and at the same time pay more attention to is when the refrigerant oil is sent into the condenser, and then through the drying and filtering device into the finer thin tube, it will produce a blockage of the pipe, this The situation is often referred to as the oil plug.