The VOC emission chamber used in the paint VOC release
Before testing the VOC release amount of the paint, the VOC emission chamber needs to measure the background concentration of the VOC discharge cabin. Follow the steps below to measure the background dripping of the VOC discharge cabin.
1. After cleaning, the door of the environmental test cabin is closed, the purified air is passed in, and the operation is started, and the time when the door is closed is 0.
2. The test time is 24h±1h, and the background concentration in the cabin is measured 24 hours after closing the cabin door.
3. Collect the background blank twice as required and test it.
4. The relative deviation of the two test results should be less than 20%, and the result after taking the average should meet the requirements.
The VOC emission chamber measures the VOC release amount of the paint:
For the determination of the VOC release amount of the coating, place the coating surface of the prepared test piece horizontally in the center of the cabin, so that the air flow uniformly passes through the surface of the test piece. The sample should be prevented from falling on the surface of the chamber, and the placement of the test piece should be completed within 30s. After the specimen is placed, the hatch door is quickly closed to start the test.
Test equipment: paint VOC emission chamber
Equipment model: HYV-1000
Equipment brand: Huanyi Instrument
The above is about the test method of the VOC emission chamber. For the specific parameters of the VOC emission chamber, please visit the official website of "Huanyi Instrument" and consult relevant technical personnel.