Salt spray test machine common faults and solutions are introduced

Date: Feb 20,2023 Views: 274

Salt spray testing machine, also known as brine spray testing machine, salt spray machine for short is one of the artificial environment three prevention series, mainly to simulate the ocean surrounding climate caused by the destruction of products, suitable for galvanized, coating, electrical, zipper buttons and other industrial products, very wide use. So in the process of use if some small problems lead to the interruption of the test and can not be dealt with in time, resulting in the need to re-test, it is very time consuming.

 

In the process of using the salt spray test machine, some common problems are actually we can deal with immediately, here is to talk about the salt spray machine common four problems and treatment.

One, can not spray

During the use of the salt spray machine suddenly does not spray, this should be experienced by many users, don't panic, we only need to do two steps to check:

1. Check whether the nozzles are blocked or broken; if any, replace the nozzles (always have one); Check whether the nozzle position is correct. If it is crooked, re-align the nozzle; Check whether the brine drum and pressure drum are short of water.

2. Check the gas path. If there is no gas at the nozzle, check whether the pressure regulating valve and solenoid valve of the gas path are bad.

Two, can not be heated

Salt spray testing machine heating system is: power -- switch -- heating tube ---- temperature control, which is divided into two heating systems, one for the pressure barrel heating system, two for the laboratory heating system. According to domestic standards, the temperature of the pressure barrel is set at 47°, and the temperature of the laboratory is set at 35°.

1. The pressure barrel cannot be heated;

If the pressure barrel cannot be heated, generally rule out whether the relay is broken (there is a small green light on the relay, which will be steady on in the heating state). If there is no problem, then measure whether there is voltage (220V) at both ends of the heating tube of the pressure barrel. After excluding these two problems, there are only two reasons: one is that the temperature sensor wire of the pressure barrel is burned out, the other is that the temperature control meter is broken. Usually replace the temperature sensor wires first (cheaper) and then replace the temperature gauge later.

Note: The heating pipes will not burn out after use for several years. If there is no improvement after replacing the temperature control meter, replace the heating pipes first.

2. The laboratory cannot be heated

The treatment method is the same as that of the pressure barrel, but the temperature control in the laboratory has a primary way through the temperature protection switch. The check should first check whether the temperature protection switch has formed a circuit (the voltmeter is adjusted to the 750V AC range, one end measures the relay, the other end measures the temperature protection place, if there is 220V voltage, it is OK, and the other end needs to replace the temperature protection).

3. High and low water levels are not displayed

The display of high and low water level is controlled by the upper and lower two float switches in the pressure barrel. First check whether there is current to the float switch. After eliminating the line problem, it can be judged that the float switch itself is broken and can be replaced.

Iv. The pressure gauge has no display or the display is large after aeration

Generally, the pressure regulating valve is broken and can be replaced. In addition, it is recommended that customers install filter cups before the gas pipe enters the pressure regulating valve. Because there is water in the air, the pressure regulating valve and relay are easy to be blocked and blocked.




Prev New Next New

Contact us

Telphone: +86 0769-83482055 / +8615338361334

Fax: +86 0769-83482056

Email: dg@huanyi-group.com

Add: Building 2, No. 3, Longkeng Xingye Road, Dongkeng Town, Dongguan,China

google+ twitter facebook printerest linkedin
Copyright 2018 by Dongguan Huanyi Instrument Technology Co., Ltd. All rights reserved. Powered by Onepound
Online Service
Top