Placement of 30m³ and 3m³ air cleaner test chamber
Placement of air cleaner test chamber:
1. 30m3 air cleaner test chamber
Central position: ground type (on the ground), desktop type (on 700mm tabletop), wall mounted type (1800 mm from lower edge to ground), ceiling type (on 700mm tabletop).
If not specified, the placement type is classified according to the height of the air outlet: place the test prototype on the tabletop if the height of the air outlet is less than 700 mm; place the test prototype on the ground if the height of the air outlet is more than or equal to 700 mm.
Note: If the cleaning function is the auxiliary function, for air conditioner, the dehumidifier, the fresh air unit etc., it shall be tested in complete-product level, however, the relevant components with the cleaning properties are only required to be started, and other components are not required to be started.
2. 3m3 air cleaner test chamber
Place the test prototype on the 400mm high tabletop if the height of the air outlet is less than 400 mm; place the test prototype on the ground if the height of the air outlet is more than or equal to 400 mm.