Introduction of Medical Mask Synthetic Blood Penetration Tester
Main Purpose:
The medical mask synthetic blood penetration tester is used to measure the penetration performance of liquid-derived liquids such as synthetic blood to evaluate its safety protection performance.
Standard:
GB19083-2010: Medical protective mask technology;YY 0469-2011: surgical masks;
YY / T0691-2008: Test method for resistance to synthetic blood penetration of medical masks for infectious pathogen protective equipment (fixed volume, horizontal spray)
EN 14683-2019 Medical face masks - Requirements and test methods
ASTM F2100-19 Standard Specification for Performance of Materials Used in Medical Face Masks
ASTM F1862 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
Test principle:
The medical mask sample was supported on the test device, and 2 mL of synthetic blood was sprayed from a cannula with an inner diameter of 0.84 mm in a horizontal direction toward the mask to be tested at a distance of 300 mm. Simulate the scene where the mask is sprayed with blood from the perforated blood vessels, and observe the penetration of synthetic blood on the other side of the mask.
The medical mask samples were tested at the liquid ejection speeds of 450 cm/s, 550 cm/s, and 635cm/s, respectively. The test results of each spray speed were recorded, and the highest blood pressure corresponding to the acceptable quality level limit of the medical mask was 4 corresponding.
Dongguan Huanyi Instruments Co.,Ltd