Installation safety information of high and low temperature test chamber
Installation safety information of high and low temperature test chamber
1. high and low temperature test chamber Power
The high and low temperature test box uses three-phase five-wire power supply, and the equipment cable is four-core wires, three phase wires, and one neutral wire. Please distinguish between zero line. Because part of the equipment load and control system are single-phase, connecting the wrong neutral wire will cause the equipment to not work and be damaged.
The grounding wire of this equipment is equipped with a grounding pole. Ensure that the equipment is well grounded to make the equipment safe and reliable. Wrong connection of the neutral wire and ground wire will cause the equipment to work abnormally.
Due to user site and other factors, the cable needs to be extended. Please ensure that the rated parameters of the extension cable and power supply used can meet the power requirements of the product.
2. high and low temperature test chamber Venue
This equipment generates a lot of heat during use. Please install the device in a well-ventilated place.
Do not use this equipment near flammable materials or in explosive environments.
Do not use in a strong magnetic field and strong vibration environment.
Use safety information
Do not connect the power supply of the sample to the power supply of the equipment, otherwise it will increase the power load of the equipment. Unless the equipment has a power interface for the test product.
It is strictly prohibited to test flammable, explosive, highly corrosive, and strong radiation items.