How to solve the sudden failure of the constant temperature and humidity test chamber?
1) constant temperature and humidity test chamber First check
Start by checking the under-water over-temperature protector in the control room to see if it is set at 150°C. If it is not set at 150°C, please turn the under-water over-temperature protector to 150°C;
2) constant temperature and humidity test chamber Check again to check whether the pipeline in the machine room is too dirty, causing the water flow to be irregular
Solution: Please turn off the main power at the rear of the machine first, and then open the drain valve under the humidification barrel to a position parallel to the pipeline. When the water in the upper water tank is all drained, use a wrench to open the humidification barrel to remove the internal scale due to long-term humidity, then unplug the silicone tube and scrub it in water. After maintenance, follow the removal method and reverse the direction. Put it back. After the installation is complete, close the drain valve under the humidification barrel to a position perpendicular to the pipeline, and then turn on the main power supply at the back of the machine.
3) constant temperature and humidity test chamber Check if there is water in the upper water tank, if there is no water, check if there is water in the lower water tank
Solution: There is no water in the upper and lower water tanks, please add water to the lower water tank to solve the problem. If there is no water in the upper water tank and there is water in the lower water tank, the water pump motor of the lower water tank is broken. Please notify the maintenance customer service department of Guangdong Kebao Testing Equipment Co., Ltd. for on-site repair.