Phone : +86 0769-83482055 / +8615338361334 Email: dg@huanyi-group.com
facebook twitter linkedin google+ printerest
You are here: Home > Technical Literature > How to set the program setting of the programmable constant temperature and humidity test chamber

How to set the program setting of the programmable constant temperature and humidity test chamber

Date: Jan 05,2022 Views: 320

Before the programmable constant temperature and humidity test chamber is tested, the program needs to be set up before the test can be entered. How to set the program? The following editor gives a simple setting method.

 

The program settings of the programmable constant temperature and humidity test chamber are as follows:

 

1. If you need to set the value to -20, work for 5 hours. Click "Catalogue" (enter the main directory of the temperature and humidity control table), then click "Operation Settings"?? Select the operating mode: fixed value

 

2. Click to switch "Next Page", running time (H) 5HOUR; running time (M) 40MIN (this time is mainly used for the cooling process); timing mode selects "ON". Click "Catalogue" again (return to the main list of temperature and humidity control tables).

 

3. Click "monitor display" (set temperature and humidity into the meter): temperature setting "-20.00", humidity setting "0.0". Click "Run" and select "Yes". Then the above is the setting method of fixed value test.

 

4. If you need to program from -20°C (3H) to +60°C (3H) to 60°C, 75% (3H), cycle 3 times. Click "Catalogue" (enter the main directory of the temperature and humidity control table), then click "Operation Settings", and select the operating mode as: Program.

 

5. Then click to switch the "next page" timing mode and select "OFF". Click "Catalogue" again (return to the main list of temperature and humidity control tables). ? Click "Program Settings" (enter the program operation directory). Click "Program Edit", click "Program" and click "Cycle" to cycle all (3).

 

 6. Finally click "Program", click "List", click "Monitor Display", click "Run", and select "Yes". ? The above is the setting method of the program test. (Note: The number of segments 01, 03, 05 is the fastest rate constant for heating and cooling).

 

The above steps can be operated by referring to the controller. If the operation is not possible, the controller also has a manual, and there are also detailed program setting operation steps. You can also read the steps in the manual, or call us directly.




Prev New Next New

Contact us

Telphone: +86 0769-83482055 / +8615338361334

Fax: +86 0769-83482056

Email: dg@huanyi-group.com

Add: Building 2, No. 3, Longkeng Xingye Road, Dongkeng Town, Dongguan,China

google+ twitter facebook printerest linkedin
Copyright 2018 by Dongguan Huanyi Instrument Technology Co., Ltd. All rights reserved. Powered by Onepound
Online Service
Top