How to deal with the refrigerant leakage of the constant temperature and humidity test chamber?

Date: Feb 10,2022 Views: 305

How to deal with the refrigerant leakage of the constant temperature and humidity test chamber? The refrigeration principle of the constant temperature and humidity test chamber is as follows: the compressor of the constant temperature and humidity chamber inhales low temperature and low pressure refrigerant gas from the suction pipe, drives the piston to compress it through the operation of the motor, and discharges the high temperature and high pressure refrigerant to the exhaust pipe. gas, which powers the refrigeration cycle, thereby a refrigeration cycle of compression--condensation--expansion--evaporation (heat absorption) is realized. What should I do if the refrigerant leaks during the test?

 

First, observe the pressure gauge on the back of the constant temperature and humidity test chamber to see if the pressure is within the normal range. If it is lower than the normal value, it means that the refrigerant is leaking. The refrigeration system of the high and low temperature test chamber should be checked for leaks. Enter high-pressure nitrogen into the tube, use a combination of leak detector and soapy water to check for leaks. Usually, there is only one leak, but sometimes there are multiple leaks, which are relatively rare.

 

After finding the leak, seal the leak with oxygen welding, then fill the refrigeration system with nitrogen, and keep the pressure for 48 hours. During the pressure maintaining process, it is found that the pointer of the pressure gauge has not changed, indicating that the leak has been repaired and welded normally, and nitrogen is released. Fill the system with environmentally friendly refrigerants R404 and R23, and the refrigeration system can resume normal operation.The BYH series high and low temperature test chambers and standard constant temperature and humidity test chambers produced by Shanghai Juyin are all cooled by air-cooled compressors.

 

The working principle of the constant temperature and humidity test chamber is the same as the principle of air conditioning refrigeration. During the cooling process, refrigerant will be used for cooling. Once the refrigerant leaks, the test chamber will stop running, low temperature test and damp heat test with humidity less than 85%RH cannot be performed.




Prev New Next New

Contact us

Telphone: +86 0769-83482055 / +8615338361334

Fax: +86 0769-83482056

Email: dg@huanyi-group.com

Add: Building 2, No. 3, Longkeng Xingye Road, Dongkeng Town, Dongguan,China

google+ twitter facebook printerest linkedin
Copyright 2018 by Dongguan Huanyi Instrument Technology Co., Ltd. All rights reserved. Powered by Onepound
Online Service
Top