1 cubic meter VOC emission environmental test chamber
Application range: VOC Emission Chamber can provide standard testing environment (such as appointed temperature, relative humidity, air exchange rate and airflow rate) or s mulate real indoor environment, and test the pollutant release rate and release prope ties of various indoor materials and products. Then it can: A.Confirm the influence of environment variables on the release rate and release properties of materials and products; B.Grade materials and products according to their release properties; C.Develop and validate the proposed model of indoor air pollutio ; D.Evaluate indoor air quality (IAQ) according to the da a and proposed model of indoor pollutants;
Sample limitations:flammable, explosive, corrosive, volatile substances, and strong electromagnetic emitters testing and stock (except for special requirements)
- Product Details
1. Product use
The equipment is mainly used for the determination of volatile organic compounds (VOC) and aldehydes and ketones in consumer products such as wood products, furniture, floors, carpets, footwear products, and building decoration materials.
2. Implementation standards
EN 717-1 "Environmental Chamber Method for Measurement of Formaldehyde Emission from Wood-based Panels"
ASTM D6007-02 "Standard Test Method for Determination of Formaldehyde Concentration in Emitted Gases from Wood Products in Small-Scale Environmental Chambers
ASTM D5116-97 Standard Guide for small scale environmental chamber determination of organic release from indoor materials and products
ASTM d6330-98 "determination of VOC (formaldehyde removal) emission in wood plate by small environmental box under specified test conditions
Standard for release
EN13419-1 "Determination of VOCs Release from Construction Products Part 1 Method for Release Test Environmental Chamber
ISO 16000-9-2006 "Indoor air. Part 9: Determination of volatile organic compounds released from construction products and furniture. Release laboratory method
ANSI/BIFMA M7.1-2011 "Standard Test Methods for VOC Emission Rates from Office Furniture Systems, Components, and Seating
3. Technical indicators
Serial number |
Technical parameters |
|
1 |
Temperature |
Adjustment range: (10~70)℃ (normal operation) 250℃ when thermally decomposed at high temperature |
Tolerance: ±0.5℃ |
||
Uniformity: 1.0℃ |
||
Volatility: ±0.3℃ |
||
2 |
Humidity |
Adjustment range: (30~80)%RH |
Tolerance: ±3%RH |
||
Uniformity: 2%RH |
||
Volatility: ±3%RH |
||
3 |
Test cabin background concentration and supply air concentration |
TVOC≤50 μg/m3(TVOC background concentration ≤ 0.04ppm) |
VOC≤5 μg/m3(background concentration of single VOC ≤ 0.004ppm) |
||
Formaldehyde ≤6 μg/m3(formaldehyde ≤ 0.005ppm) |
||
4 |
Air replacement rate |
Adjustment range: 0.2~2m3/h (ventilation rate 0.2~2 times/hour) |
Tolerance: ±3% |
||
5 |
Differential pressure control |
Adopt imported differential pressure transmitter and bus transmission technology, which improves the measurement accuracy, and its accuracy can reach ±1Pa. Matching the differential pressure control device, it can automatically adjust the relative pressure difference inside and outside the cabin according to the change of flow. |
6 |
Airtightness |
The air leakage per minute is less than 1% of the test chamber volume at a pressure of 1000PA.; The difference between the air inlet and outlet is not more than 2%; |
7 |
Air Tissue |
Design the windshield and pipeline according to the principle of aerodynamics, install the air guide groove behind the windshield to ensure uniform air supply; adopt multi-point frequency conversion speed regulation technology to ensure that the air in the cabin is evenly mixed and its uniformity It satisfies that there is no condensation phenomenon when the cabin temperature is at the lowest design point. Air velocity in the cabin (0.1~0.5)m/s |
Display accuracy: ±0.01m/s |
||
Control accuracy: 0.05m/s |
||
8 |
Recovery rate |
The recovery rate of toluene or n-dodecane is more than 80% |
9 |
Structure |
Jacketed structure: test volume of the environmental chamber tank (1±0.01) m3 |
10 |
Inner bladder and gas path Material |
The inner wall of the test chamber, pipes and accessories (such as valves, joints, circulating fans, etc.) is made of mirror stainless steel; sealing material: non-polluting, non-adsorbing materials are used. |
11 |
Insulation box |
White coating on the outer wall; inner wall: 304 stainless steel plate, rock wool insulation, excellent performance |
12 |
High temperature cleaning system for test chamber |
This equipment is equipped with a high-temperature cleaning system. It has a high-temperature cleaning system. It uses high-efficiency electric heating equipment to heat the inner wall of the test chamber. The high-temperature cleaning temperature can reach 250 °C. The thermal efficiency is high and the energy consumption is low. produce environmental pollution. It has a complete monitoring and alarm device, which monitors the running status in real time during the entire high-temperature cleaning process, and has heater overheating, overcurrent, and overheating and overpressure protection alarms in the cabin to avoid damage to the equipment during the high-temperature cleaning process. The test chamber adopts the one-time forming technology of thermal welding. After welding, high temperature treatment is performed to control the deformation of the chamber body within a certain range during high temperature cleaning. At the same time, our company's unique elastic connection technology is used to make the heating process The thermal deformation of the medium is eliminated in time after cooling, which effectively prolongs the service life of the equipment. |
13 |
Measurement and control system
|
Using high-resolution intelligent touch screen controller and Siemens PLC, the temperature, humidity, dew point temperature, working time, gas flow and other parameters of the climate chamber can be set through the touch screen; the operation process of the equipment is simple and intuitive, and the experiment can be carried out according to the prompts. With U port, data can be exported. It can be monitored by computer terminal or mobile APP. |
14 |
Noise Control |
Under normal operating conditions, less than 60dB |
15 |
Dimensions |
W1230mm×H1670mm×D2150mm |
16 |
Inner tank size |
W800mm×H1098mm×D1140mm |
17 |
Power |
Maximum 14KW |
18 |
Humidity control method |
Combination of dew point method and dry and wet air mixing method |
19 |
Continuous working hours |
≥30 days |
20 |
Insulation resistance |
The insulation resistance of the live part of the climate box to the box shell is not less than 1 MΩ, and the box body is well grounded and has obvious signs. |
21 |
Withstand voltage |
When the terminal (including the lead wire) of the climate box heater and the control system are open-circuited, the box can withstand the voltage 1500 V, AC 50 Hz, 1 minute withstand voltage strength test, and its insulation has no breakdown or flashover phenomenon. |
22 |
Other sizes |
Specific volume can be customized according to requirements (0.02-1m3) |