1 cubic meter metal corrosion aerosol cabinet
Product use:
Determine whether the sterilized equipment and the dose of sterilization are corroded
- Product Details
Meet the standard:
GB/T27948-2020 General requirements for air disinfectants
Technical parameters:
Test chamber volume |
1 cubic meter metal corrosion aerosol cabinet |
Test cabin dimensions |
800 × 1000 × 1250mm (L×W×H), allow ±0.1m3 deviation |
Temperature range |
23℃±2℃, adjustable |
Humidity range |
50%±10%R Adjustable |
Temperature fluctuation |
≤±0.5℃ |
Evenness |
Temperature ≤1℃; (no load) |
Relative Humidity Fluctuation |
±2% |
Cabin |
Stainless steel cabin, the inner and outer walls are made of SUS304 stainless steel with a thickness of 1.0 mm; the middle is high-density PU foam, which has a good thermal insulation effect. |
Inner bladder |
The inner tank material is made of wide-format mirror 304 stainless steel, which is fully welded. |
Door |
The size of the door is 1000×1250mm (W×H), single-opening is fully sealed, and the sealing is made of PTFE strip sealant (Teflon) (HBC1079 * grade sealant has passed the EU ROHS environmental protection standard, passed the EU REACH chemical regulation standard, Through the US FDA food hygiene standards). |
Sealing material |
With silicone rubber strip and glass sealant |
Test phase |
1. Put the used equipment into the aerosol cabinet (or room) at one time, and close the door. After that, all operations and manipulations of instruments and equipment are carried out in the cabinet (or outdoors through a window or remote control with a sealed sleeve. The door cannot be opened until the end of the test. 2. The spray device and the disinfection machine are operated according to each cycle time, starting, spraying, sterilizing, and ventilating for 45 operations. 3. At the same time, the aerosol cabinets (or chambers) of the control group and the test group were sampled before disinfection, and the control group used the test water for repeated tests. 4. After the test, carry out cleaning and weighing to calculate the corrosion rate. |