Phone : +86 0769-83482055 / +8615338361334 Email: dg@huanyi-group.com
facebook twitter linkedin google+ printerest
You are here: Home > Technical Literature > Air disinfectant sterilization performance bacterial aerosol chamber sterilization requirements

Air disinfectant sterilization performance bacterial aerosol chamber sterilization requirements

Date: Mar 26,2022 Views: 300

The sterilization test of air disinfectants requires the use of bacterial aerosol chamber. The sterilization requirements of disinfectants are divided into laboratory sterilization tests and field tests. Let's take a look.

 

Air disinfectant sterilization requirements:

 

Equipment used: 1m/20m bacterial aerosol chamber

 

Device model: HYKJ

 

Equipment manufacturer: Huanyi Instrument

 

1. Laboratory sterilization test: under the conditions of 20 ~ 25 and relative humidity of 50% ~ 70%, the effect of the disinfectant is 1h, and the killing rate of Staphylococcus albus (8032) in the air should be 99.90%. When using the aerosol spray method for disinfection, the amount of disinfectant should be less than or equal to 10 mL/m.

 

2. Field test: Under natural conditions, the effect of disinfectant is less than or equal to 1h, and the extinction rate of natural bacteria in the air should be more than 90.0%. When using the aerosol spray method for disinfection, the amount of disinfectant should be less than or equal to 10 mL/m.

 

Air Sanitizer Sterilization Test Bacteria Aerosol Chamber Picture:





Prev New Next New

Contact us

Telphone: +86 0769-83482055 / +8615338361334

Fax: +86 0769-83482056

Email: dg@huanyi-group.com

Add: Building 2, No. 3, Longkeng Xingye Road, Dongkeng Town, Dongguan,China

google+ twitter facebook printerest linkedin
Copyright 2018 by Dongguan Huanyi Instrument Technology Co., Ltd. All rights reserved. Powered by Onepound
Online Service
Top