Air disinfectant sterilization performance bacterial aerosol chamber sterilization requirements
The sterilization test of air disinfectants requires the use of bacterial aerosol chamber. The sterilization requirements of disinfectants are divided into laboratory sterilization tests and field tests. Let's take a look.
Air disinfectant sterilization requirements:
Equipment used: 1m/20m bacterial aerosol chamber
Device model: HYKJ
Equipment manufacturer: Huanyi Instrument
1. Laboratory sterilization test: under the conditions of 20 ℃ ~ 25 ℃ and relative humidity of 50% ~ 70%, the effect of the disinfectant is ≤ 1h, and the killing rate of Staphylococcus albus (8032) in the air should be ≥ 99.90%. When using the aerosol spray method for disinfection, the amount of disinfectant should be less than or equal to 10 mL/m.
2. Field test: Under natural conditions, the effect of disinfectant is less than or equal to 1h, and the extinction rate of natural bacteria in the air should be more than 90.0%. When using the aerosol spray method for disinfection, the amount of disinfectant should be less than or equal to 10 mL/m.