12m³ VOC Emission Chamber
12 cubic meters of VOC release detection environmental chamber. Benzene, toluene, and other components in volatile organic compounds (VOCs) are recognized by the World Health Organization as strong carcinogens. Long term exposure to pollutants with excessive concentrations can seriously harm people's health. Testing and limiting the release of pollutants will be beneficial for improving product quality and protecting people's health.
Wood products, furniture, flooring, carpets, footwear products, building decoration materials, automotive interior materials, and other products that come into contact with the human body can emit harmful substances such as volatile organic compounds (VOC) and formaldehyde, especially for indoor and enclosed spaces where products are densely placed, the concentration will be higher after exposure, posing greater harm to human health.Product introduction: The applicable standard for the 12 cubic meter large environmental test chamber is ISO 12219-2-2012 Determination of Volatile organic compound Emission from Automotive Interior Trim Materials by Sieving -- Bag Method.- Product Details
The 12 cubic meter VOC emission detection environmental chamber is mainly used to detect formaldehyde and VOC emissions in interior decoration materials, decoration materials, artificial boards, and their wood products; At the same time, other products can also be tested for VOC release. Comprehensive environmental indicator evaluation of indoor air quality based on indoor pollutant data or prediction models.
Meet standards:
GB 18580-2001 Indoor Decoration Materials - Limits of Formaldehyde Emission from Engineered wood and Their Products GB 18584-2011 Indoor Decoration Materials - Limits of Harmful Substances in Wood Furniture (Appendix A (normative appendix) of the approved GB 18587-2001 Indoor Decoration Materials - Limits of Harmful Substances Emission from Carpets, Carpet Villages and Carpet Agents): small environmental test chamber method.
GB 50325-2010 Code for Indoor Waste minimisation of Civil Building Engineering
GB/T 18584 Climatic General-purpose technology Conditions for the Detection of Volatile organic compound in Furniture (Draft for Approval) GB/T 284-2012 Limits of Harmful Substances in Musical Instruments
ILY/T 1980-2011 Volatile organic compound and Formaldehyde Emission Test Chamber
ILY/T 1981-2011 "Detection Box for Formaldehyde Release Gas Analysis Method"
UG/T 344-2011 Indoor Environment Test Tire for Building Engineering
IHJ 571-2010 (Technical Requirements for Environmental Labeling Products Engineered wood and Products)
IHJ/T 201-2005 "Technical Requirements for Environmental Labeling Products - Waterborne Coatings"
IHJ/T 220-2005 "Technical Requirements for Environmental Labeling Products - Adhesives"
IHJ/T 414-2007 "Technical Requirements for Environmental Labeling Products - Solvent based Wood Coatings for Indoor Decoration and Decoration"
Specification
Model
HYVW-12m³
Ambient temperature
5℃-35℃
Ambient humidity
≤90R.H
Environmental cleanliness requirements
No pollution sources around
Climate box volume
3*2*2m(W*D*H)
Door
1000*1800
Temperature range
20-100℃ High temperature cleaning 250℃
Humidity range
40-70R.H.(20-30℃),According to VDA276, the temperature is 56 (at 65 ℃)
Pressure
10±5pa
Temperature fluctuation
±0.3℃
Temperature deviation
±0.5℃
Humidity fluctuation
±2.5R.H.
Humidity deviation
±30R.H.
Air exchange rate
0.2-5times/h(Optional with larger exchange rate specifications)
Power Supply
AC380V/50Hz/Three-phase power supply 4/5-wire system
Starting power
20kw
Operating power
9kw
VOC background concentration
Single VOC concentration less than 2ug/m3 Dust filtration TVOC concentration less than 20ug/m
Rate of recovery
The recovery rate of formaldehyde, octanol, 4-phenylcycloethylene, dodecane, Tetradecane is more than 80%
Clean air system
Standard configuration